82
708-DS / 709-DS Operator’s Manual
4
Operating the 708-DS / 709-DS
Menu Screen - Instrument
Menu Screen - Instrument Settings Screen 1
Table 6
Menu Screen - Instrument Settings Screen 1
Option
Function
Tester ID
A twelve-character field that helps identify the instrument. The Tester
ID is printed out with the other system information during a test.
Comm ID
The address of the instrument on the communications bus. This address
is used by external host programs to communicate with the instrument.
The Comm ID may range from 01 to 99.
All Position Spin
Allows the spindle to rotate regardless of the position of the drive unit.
This is useful when calibrating spindle speed as the spindle may rotate
while the drive unit is in its upper position. If this option is not checked,
the spindle only rotates when the drive unit is in its lowest position.
Содержание 708-DS
Страница 1: ...Agilent Technologies 708 DS 709 DS Dissolution Apparatus Operator s Manual ...
Страница 10: ...10 708 DS 709 DS Operator s Manual Tables This page was intentionally left blank except for this message ...
Страница 125: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 125 Figure 29 Resistance Value of Temperature Probe ...
Страница 133: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 133 Figure 37 Feeding Paper ...
Страница 136: ...136 708 DS 709 DS Operator s Manual 5 Maintenance and Troubleshooting ...
Страница 140: ...138 708 DS 709 DS Operator s Manual Index This page was intentionally left blank except for this message ...