![Agilent Technologies 708-DS Скачать руководство пользователя страница 112](http://html1.mh-extra.com/html/agilent-technologies/708-ds/708-ds_operators-manual_2867627112.webp)
112
708-DS / 709-DS Operator’s Manual
4
Operating the 708-DS / 709-DS
Spindle Speed Error
This Spindle Error screen displays when the spindle speed has exceeded
the allowable tolerance range (
“Menu Screen - Instrument Settings
on page 84). Press
Ok
to resume the test and send a message
to the printer, if installed.
Fatal Spindle Error
•
This Spindle Error screen displays when a serious error has
been detected with the spindle motor. For the safety of the
equipment and the operator, remove power before
correcting the problem. Errors may include jammed
paddles, obstructions in the path of the drive belt, and
severe deterioration of the spindle motor.
•
Contact Agilent Customer Care (
Autosampling Manifold Errors
The Stepper Motor Errors screen shows all of the possible errors
associated with the manifold stepper motor. A red X displays before the
current error condition. If a stepper motor error is encountered, the
current run is stopped. You can press
Ignore
to disregard the error, but it
must be corrected before a run may be restarted. Press
Retry
to re-test
the stepper motor for errors.
Autosampling Manifold Error
Action
Over Current
•
Contact Agilent Customer Care (
Over Temperature
Under Voltage
Controller Command Retry Error
Open Load
Timeout
Table 17
Error Screens (Red)
Содержание 708-DS
Страница 1: ...Agilent Technologies 708 DS 709 DS Dissolution Apparatus Operator s Manual ...
Страница 10: ...10 708 DS 709 DS Operator s Manual Tables This page was intentionally left blank except for this message ...
Страница 125: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 125 Figure 29 Resistance Value of Temperature Probe ...
Страница 133: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 133 Figure 37 Feeding Paper ...
Страница 136: ...136 708 DS 709 DS Operator s Manual 5 Maintenance and Troubleshooting ...
Страница 140: ...138 708 DS 709 DS Operator s Manual Index This page was intentionally left blank except for this message ...