Installing 7890 GC Columns
To Calibrate the Column
7000/7010 Series TQ GC/MS Operating Manual
93
To Calibrate the Column
Capillary columns must be calibrated prior to use with the MS. This procedure
applies only to the 7890 GC.
Procedure
1
Set Data Acquisition for splitless manual injection and selected ion
monitoring (SIM) of
m/z
28.
2
Press [
Prep Run]
on the GC keypad.
3
Inject 1 µL of air into the GC inlet and press [
Start Run
].
4
Wait until a peak elutes at
m/z
28. Note the retention time.
5
In the
Instrument Control
panel, select
Instrument > GC Configuration
.
6
Select the
Configuration
tab.
7
Select the
Column
tab, click
Inventory
, and verify that the column you are
using is in inventory. Select the column to be calibrated and click
Install
Selected Column
.
8
Highlight the column in the inventory list and select
Calibrate
.
9
Select
Calc Length
.
10
In the
Calculate Column Length
dialog, enter the recorded retention time in
the
Holdup Time
field. Verify that the other parameters listed (temperature,
inlet and outlet pressures, and gas type) are those used in the method to
determine the holdup time. Change any parameters that are different than
those used in your method. (See
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...