3
Installing Intuvo 9000 GC Column
To Install a Column Guard or Jumper Chip
58
7000/7010 Series TQ GC/MS Operating Manual
To Install a Column Guard or Jumper Chip
The column Guard Chip and the Jumper Chip are both single-use, consumable
parts. Installation deforms part of the Chip to make a good seal, so that a
mis-installed Chip cannot be re- used. The Guard Chip cannot be cleaned or
conditioned.
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Tweezers (8710-2460)
•
Split/Splitless inlet Intuvo guard chip, 2/pk (G4587-60565)
•
Multimode inlet Intuvo guard chip, 2/pk (G4587-60665)
•
Split/Splitless inlet Intuvo jumper chip, 2/pk G4587-60575)
•
Multimode Inlet Intuvo jumper chip, 2/pk (G4587-60675)
•
Intuvo gasket, polyimide 5/pk, for temperatures < 350 °C (5190-9072)
•
Intuvo gasket, nickel 5/pk, for temperatures 350 °C to 450 °C (5190-9073)
•
Intuvo torque driver stored on oven door (5190-9571)
•
7/16-inch open-end wrench
Procedure
1
Prepare the GC for maintenance. From the GC panel select
Maintenance >
Inlets > Guard Chip > Prepare for Maintenance >Replace Liner and Guard
Chip > Start Maintenance
. This procedure cools the inlet, detector, column,
Guard Chip, and other components in the flow path heated zones to < 40 °C
and configures the GC. Follow the on-screen GC prompts.
2
From MassHunter Data Acquisition, Vent the MS (See
WARNING
The inlet, detector, bus components, and column can be hot enough to cause
burns. Cool heated zones to a safe handling temperature before continuing.
CAUTION
Always wear clean gloves while handling any parts that are in the flow path
going to the analyzer or parts that are located inside the analyzer.
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...