Installing Intuvo 9000 GC Column
To Replace an Intuvo GC Column
7000/7010 Series TQ GC/MS Operating Manual
55
8
Unplug the column SmartID key from the lower USB port.
9
Using the Intuvo torque driver, remove the two compression bolts that seal
the column's click and run connectors to the bus and store for later use.
10
Remove and store the column for later use in accordance with the column
manufacturer's recommendations.
11
Replace the gasket with a new gasket rated for the maximum column
temperature expected with your method. (See
Figure 10.
9000 GC column and related parts
SmartID key
Column clamps
Retainer ring
Compression
bolts
Click and run
connectors