
8
CI Maintenance
To Switch from the CI Source to the EI HES
7000/7010 Series TQ GC/MS Operating Manual
213
To Switch from the CI Source to the EI HES
Procedure
1
Vent the MS. (See
on page 124.) The software will prompt
you for the appropriate actions.
2
Open the left side access panel. (See
“To Open Left Side Panel for Access to
3
Open the front analyzer chamber. (See
4
Remove the CI source. (See
“To Remove the 7000 Series CI Source”
5
Place the CI source in the storage container.
6
Remove the CI source radiator. The Radiator should not be removed or
installed with the CI source in place. (See
Radiator from the 7010 Series TQ”
7
Place the CI source radiator in the storage container.
8
Loosen the column nut and remove the column from the GC/MS Interface.
9
Cut the column on the tapered side of the ferrule to remove the ferrule.
10
Install the column into the GC/MS interface. (See
Column in the GC/MS Interface Using the Self-Tightening Column Nut”
page 37 or
“To Install a Capillary Column in the GC/MS Interface Using a
11
Remove the EI HES radiator from the storage container.
12
Install the EI HES radiator. (See
“To Install the EI HES Radiator”
13
Remove the EI HES from the storage container.
14
Install the EI HES. (See
15
Pumpdown the MS. (See
CAUTION
Always wear clean gloves while touching the analyzer or any other parts that
go inside the analyzer chamber.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap and take other antistatic precautions
before
you open the
analyzer chamber. (See
Содержание 7010 Series
Страница 1: ...Agilent 7000 7010 Series Triple Quadrupole Operating Manual ...
Страница 12: ...12 7000 7010 Series TQ GC MS Operating Manual To Install a CI Source Filament 240 ...
Страница 128: ...5 Operating in EI Mode To Save a Method on the 7890 GC 128 7000 7010 Series TQ GC MS Operating Manual ...
Страница 204: ...7 General Maintenance To Close the Rear Analyzer Chamber 204 7000 7010 Series TQ GC MS Operating Manual ...
Страница 241: ......
Страница 242: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7003 90050 G7003 90050 ...