CI Troubleshooting
3
Troubleshooting and Maintenance Manual
57
Figure 3
Schematic of CI flow control module
Gas A
(methane)
supply
Gas B
(other gas)
supply
Gas A
select valve
Gas B
select valve
Mass
flow
controller
Calibration
valve
Restrictor
Calibration
vial
Shutoff valve
GC/MS
interface
GC column
CI ion
source
Table 4
Flow module valve state diagram
Result
Gas A flow
Gas B flow
Purge
with Gas A
Purge
with Gas B
Pump out
flow module
Standby,
vented, or
EI mode
Gas A
Open
Closed
Open
Closed
Closed
Closed
Gas B
Closed
Open
Closed
Open
Closed
Closed
MFC
On (at
setpoint)
On (at
setpoint)
On (at 100%)
On (at 100%)
On (at 100%)
Off (at 0%)
Shutoff valve
Open
Open
Open
Open
Open
Closed
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...