Analyzers and Collision Cell
7
Troubleshooting and Maintenance Manual
163
Analyzer Heaters and Radiators
The ion source and mass filter are housed in cylindrical aluminum tubes
called radiators (
). The radiators control the distribution of heat in
the analyzer. They also provide electrical shielding for analyzer components.
The source heater and temperature sensor are mounted in the source heater
block. The mass filter (quad) heaters and temperature sensors are mounted on
the mass filter radiators. Analyzer temperatures can be set and monitored
from the MassHunter Workstation software.
In selecting the temperatures to use, consider the following:
•
Higher temperatures help keep the analyzer clean longer.
•
Higher ion source temperatures result in more fragmentation and therefore
lower high-mass sensitivity.
After pumpdown, it takes at least 2 hours for the analyzer to reach thermal
equilibrium. Data acquired sooner may not be reproducible.
Recommended settings (for EI operation):
•
Ion source 230
°
C
•
Quadrupoles (front and rear)150
°
C
The GC/MS interface, ion source, and front mass filter (quad) heated zones
interact. The analyzer heaters may not be able to accurately control
temperatures if the setpoint for one zone is much lower than that of an
adjacent zone.
C A U T I O N
Do not exceed 200 °C on the quadrupole or 350 °C on the ion source.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...