Analyzers and Collision Cell
7
Troubleshooting and Maintenance Manual
157
Width gain
Width gain affects the ratio of DC voltage to RF frequency on the mass filter.
This controls the widths of the mass peaks.
•
Higher gain yields narrower peaks.
•
Width gain affects peaks at high masses more than peaks at low masses.
Width offset
Width offset also affects the ratio of DC voltage to RF frequency on the mass
filter.
•
Higher offset yields narrower peaks.
•
Width offset generally affects peak widths equally at all masses.
Mass gain
Mass gain controls the mass assignment, that is, assignment of a particular
peak to the correct
m/z
value.
•
A higher gain yields higher mass assignment.
•
Mass gain affects peaks at high masses more than peaks at low masses.
Figure 30
Quadrupole mass filter
C A U T I O N
Using the nonpreferred DC polarity may result in very poor performance. Always use
the factory-specified polarity.
Содержание 7000 Triple Quadrupole GC/MS
Страница 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Страница 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Страница 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Страница 193: ...Replacement Parts 9 Troubleshooting and Maintenance Manual 193 Figure 45 Foreline pump top and safety cage parts 1 6 5 7...
Страница 212: ...212 Troubleshooting and Maintenance Manual 9 Replacement Parts Figure 50 CI flow control module 5 3 4 2 1...
Страница 213: ......
Страница 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...