Valves
Agilent 6890 Gas Chromatograph Service Manual
Valves
1110
Valco W-series minivalves
Valve bodies
Body parts are made from Nitronics 60 nickel steel. If required, the valve may
also be produced from Hastelloy C. External tubing (plumbing) is connected
to the valve body ports by ferrules and fittings provided with the instrument.
The left (CCW, counterclockwise) and right (CW, clockwise) stops on general
purpose valve bodies limit rotor rotation so the correct flow path results
when the index pin is close to or against either stop of the index lip.
Caution
Intermediate positions of the rotor may result in an interrupted flow path
which could cause damage to the valve or other components in the
chromatograph.
Valve rotors
Rotor type can be identified by color:
•
An
off-white
rotor is made of a PTFE composite and may be used from
room temperature to 200
°
C.
•
A
black
rotor is made of polyimide and may be used from 100 to 350
°
C.
Caution
The life of a valve is shortened if used outside its specified temperature range
Do not mix rotor types in the same system.
The rotor seats on a highly polished conical surface. When properly seated,
the polished surface prevents leakage around the rotor and between non-
selected ports. The finish precludes adsorption of most GC samples.
The rotor assembly is a one-piece part with an integral molded and machined
conical hub and the parts necessary for proper seating. The sample contacts
only the PTFE composite (low temperature) or polyimide (high temperature)
as well as the stainless steel of the valve.
Grooves in the rotor surface form the paths between specific ports. The index
pin prevents rotation beyond either stop of the index lip. Valve ports are
connected by the grooves only when the index pin is close to or against either
stop. Intermediate positions result in flow shutoff through the valve and
possible damage if left in this position.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......