Faults
Fault 218—F TCD filament shorted
6 of 22
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
board, or the
∆
PRT is shorted. (Units made before April 1997 may not use a
∆
PRT.)
The detector will not be ready until the condition is corrected.
❐
Check that the wires from the detector are connected on the detector
board.
❐
Check that the
∆
PRT is properly installed and not damaged. Replace if
damaged.
❐
The TCD’s cell must be replaced.
Fault 218—F TCD filament shorted
Fault 219—B TCD filament shorted
Status message: F TCD filament short or B TCD filament short
The TCD filament bridge voltage indicates that the resistance of the filament
is too low, which indicates a potentially shorted filament. This condition could
be caused by a worn or sagging filament, or if the wires from the TCD are not
connected properly to the detector board or are touching each other, or if the
wires from the
∆
PRT are not properly connected to the detector board. (Units
made before April 1997 may not use a
∆
PRT.)
The detector will not be ready until the condition is corrected.
❐
Check that the filament and
∆
PRT wires from the cell are connected on
the detector board properly.
❐
The TCD’s cell must be replaced.
Fault 220—Heater over current. Thermal shutdown
Status message: Heater overcurrent
This message appears when there is a short in the heater of a heated zone, or
when a heated zone’s electronics are defective. This error does
not
indicate
a problem with the oven.
Power to the small zones is disabled. To clear the error message, turn off all
the heated zones and turn the GC off and then on again.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......