
850
Faults
Fault 203—Signal DSP faulty
2 of 22
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
The sequence would continue until the fault is fixed. To find the fault, check
for the following possible causes:
❐
Check the gas supply pressure. Increase the pressure at the initial supply
if it is too low to reach the setpoint.
❐
Check for a leak somewhere in the system. Leak test the gas supply tubing,
the inlet, and the inlet column fittings. Leak test procedures are found
with each inlet section.
❐
The column may be broken. Use the leak detector to check the column
for leaks and replace the broken column or break off the cracked portion.
❐
An inlet proportional control valve may be stuck open or closed because
of contamination or other fault. Contact your Agilent Technologies
service representative.
Fault 203—Signal DSP faulty
Status message: Signal DSP faulty
The detector’s signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board. The signal
path will not function.
❐
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.
Fault 204—Signal DSP ROM broken
Status message: Sig DSP ROM broken
The detector’s signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board. The signal
path will not function.
❐
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......