830
Warning Codes
Warning 109—Front inlet: config changed, method defaulted
4 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning 109—Front inlet: config changed, method defaulted
Warning 110—Back inlet: config changed, method defaulted
Status message: F inl config changed or B inl config changed
This message appears when:
•
The 6890 GC is turned on after a new inlet has been installed.
•
The 6890 GC is turned on and senses that the inlet has a different type
of sensor than was there when the instrument was turned off.
•
A method is loaded that calls for a different type of inlet than the one
currently installed.
The following will occur:
❐
If a new inlet or sensor was installed, the 6890 GC will enter the default
setpoints for the new type of inlet.
❐
If a method was loaded with a different inlet specified, the 6890 GC will
not change the method but will flash the warning message.
Warning 111—Column 1: config changed, method defaulted
Warning 112—Column 2: config changed, method defaulted
Status message: Col 1 config changed or Col 2 config changed
This message appears when:
•
The 6890 GC is turned on after a different column has been installed.
•
You load a method that calls for a different type of column than the one
currently installed in the oven.
The following will occur:
❐
If a new column was installed, the 6890 GC will enter the column default
setpoints.
❐
If a method was loaded for a different type of column than the one
installed, the 6890 GC will not change the method but will flash the
warning message.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......