Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning Codes
830
Warning 105—Analog out data loss
Warning 105—Analog out data loss
Status message: Analog out data loss
The 6890 GC analog processors are not communicating with each other
correctly. Analog data will be lost; digital data should be all right.
Although the problem could be caused by a bad power supply or a radio
frequency interference, both are improbable. The most likely cause is a
malfunction on the main PC board.
Warning 106—Non-recoverable data loss. Data corrupt.
Status message: Signal data loss
The 6890 GC digital processors are not communicating with each other
correctly. Both analog and digital data will be lost.
Although the problem could be caused by a bad power supply or a radio
frequency interference, both are improbable. The most likely cause is a
malfunction on the main PC board.
Warning 107—Front det: config changed, method defaulted
Warning 108—Back det: config changed, method defaulted
Status message: F det config changed or B det config changed
This warning appears when the 6890 GC is turned on after a different detector
has been installed, or when a method is loaded that calls for a different type
of detector than the one that is currently installed.
The following will occur:
❐
If a new detector was installed, the 6890 GC will enter the default
setpoints for the new type of detector.
❐
If a method was loaded with a different detector specified, the 6890 GC
will not change the method but will flash the warning message.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......