610
Checkout Chromatograms
TCD checkout conditions and chromatogram
10 of 26
Jun 2001
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
TCD checkout conditions and chromatogram
Column and sample
Type
HP-5 30 m × 0.32 mm × 0.25 µm PN 19091J-413
Sample
FID Checkout PN 18710-60170
Injection volume
1 µL
Inlet
Temperature
250
°
C
Purged/Packed or Split/Splitless
Oven Track Cool On-Column
40°C
PTV (see below)
Inlet pressure
25 psi
(Constant pressure for EPC inlets,
helium)
Split/Splitless Only
Mode
Splitless
Purge flow
60 mL/min
Purge time
0.75 min
PTV—Splitless Mode
Inlet temperature
40°C
Initial time
0.1 min
Rate 1
720°C/min
Final temp 1
350°C
Final time 1
2 min
Rate 2
100°C/min
Final temp 2
250°C
Final time 2
0 min
Inlet pressure
25 psi (const. press. for EPC inlets)
Mode
Splitless
Purge time
0.75 min
Purge flow
60 mL/min
Detector
Temperature
300
°
C
Reference flow (He)
20 mL/min
Makeup flow (He)
2 mL/min
Offset
Should be
≤
30 display counts
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......