520
Gases
The gas plumbing
8 of 14
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Traps
Using chromatographic-grade gases insures that the gas in your system is
pure. However, for optimum sensitivity, it is highly recommended that you
install high-quality traps to remove traces of water or other contaminants.
After installing a trap, check the gas supply lines for leaks.
Table 520-4
Recommended Traps
Moisture in carrier gas damages columns. We recommend a type 5A Molecular
Sieve trap after the source regulator and before any other traps.
A hydrocarbon trap removes organics from gases. It should be placed after a
molecular sieve trap and before an oxygen trap, if they are present.
An oxygen trap removes 99% of the oxygen from a gas plus traces of water. It
should be last in a series of traps. Because trace amounts of oxygen can
damage columns and degrade ECD performance, use an oxygen trap with
carrier and ECD gases. Do not use it with FID, FPD, or NPD fuel gases.
Figure 520-3
Traps
Description
Part No.
Preconditioned moisture trap: metal casing, s-shaped trap for carrier gas cleanup.
Contains Molecular Sieve 5A, 45/60 mesh, and 1/8-inch fittings.
5060-9084
Hydrocarbon trap: metal casing, s-shaped trap filled with 40/60 mesh activated
charcoal, and 1/8-inch fittings
5060-9096
Oxygen trap (for carrier and ECD gases): metal casing, and 1/8-inch brass fittings.
Oxygen trap cannot be reconditioned.
3150-0414
Molecular sieve or hydrocarbon trap—both are S-shaped
Oxygen trap
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......