210
Split/Splitless Inlet
EPC inlet
4 of 30
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Figure 210-2
EPC control splitless mode flow diagram (pre-run to purge time).
Pulsed splitless mode
For EPC inlets set to pulsed splitless mode, the column head pressure is
increased to some selected value and selected time period prior to sample
introduction. Like the split mode, the increased flow through the column
during the injection period minimizes sample losses in glass liners.
Gas saver mode
For EPC inlets, Gas Saver mode may be used in splitless or split mode of
operation for conservation of split and inlet purge flows when not in use or
in between runs. A gas saver flow rate and start time can be set independently
of normal flow rates being used.
Flow
limiting
frit
PV1
Proportional
valve 1
Flow
sensor
Septum holder
Pressure
sensor
Septum purge
regulator (not
adjustable)
Septum purge
3 mL/min
Inlet vent (split)
PV2
Proportional
valve 2
Valve
closed
1 m
L/
m
in
1m
L/
m
in
PS
SPR
FS
Safety shutdown mode:
Proportional valve 1 closed
Proportional valve 2 open
On/off valve open
V
4 mL/min
To Detector
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......