Flame Photometric Detector (FPD)
Replacement procedures
20 of 36
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Replacing the heater/sensor assemblies
1.
Turn off power to the gas chromatograph and disconnect the main power
cord. Remove the detector covers.
WARNING
Always turn off power to the gas chromatograph and allow the heated zones
to cool. Unplug the power cord.
2.
Allow time for heated zones to cool to safe temperatures.
Caution
Always turn the electrometer or the main power off before removing the PMT
housing to avoid destroying the tube.
3.
Pull the PMT housing off the detector module and remove the filter from
the detector. Set both in a safe place. Keep the open end of the PMT
covered whenever practical to avoid light damage to the tube.
4.
Remove the exhaust tubing.
5.
Remove the sheet metal cover. On the single wavelength detector, it is
held by two screws at the top and two at the bottom, while on the dual
wavelength detector it is held by two screws at the top.
6.
If replacing the upper (FRONT) heater/sensor, remove it from the
detector module. Install a new heater/sensor. If not replacing the AUX1
heater, reassemble in reverse order.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......