Flame Photometric Detector (FPD)
Replacement procedures
6 of 36
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
7.
If removing the detector board(s), loosen the captive mounting screw in
each board bracket and remove each board from the GC.
8.
If removing the EPC flow module, disconnect the ribbon cable from the
pneumatics PCB and remove the mounting screw in the top of the module.
If not removing the EPC flow module, disconnect the gang fitting at the
module. Inspect the O-rings in the gang fitting and replace if damaged.
Straighten the tubing from the gang fitting and slide the module out of
the GC.
9.
Remove the four mounting screws in the FPD bracket and remove the
FPD.
10. Reassemble in reverse order.
Column installation
Special FPD adapters
The FPD uses special adapters for capillary columns and 1/8-inch OD PTFE
columns. The FPD Capillary Adapter allows fused silica columns as large as
530
µ
m ID to be run right to the base of the FPD flame, minimizing sample
tailing or loss of chemically active sites. The FPD 1/8-inch OD Adapter allows
installation of PTFE columns concentrically around the FPD fused silica liner,
eliminating exposed hot stainless steel.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......