Micro-cell Electron Capture Detector (
µ
-ECD)
Diagnostics
22 of 26
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Diagnostics
Frequency test
Perform this test to make sure that the base frequency for the
µ
-ECD during
a blank run indicates a relatively contaminant-free system.
Note
It may take 24 hours for the
µ
-ECD baseline to completely stabilize, especially
if you are starting with a cold system and want to assure high-sensitivity
operation.
Therefore, for the most accurate results, run the detector at normal operating
conditions for as long as possible (at least 2 hours and up to 24 hours) before
running the frequency test.
If you will be injecting into an unused inlet, you must use low-bleed septa.
Make sure to condition new septa before use in an inlet for several hours with
1 to 5 mL/min carrier flow.
1.
Make sure you are using normal operating conditions and that
at least
2 hours have elapsed since the last run.
2.
Turn on the
µ
-ECD and the corresponding signal.
3.
Check the displayed “Output”:
•
<25 =
µ
-ECD frequency is acceptable
•
≥
1000 = Contaminants in system
Note
Each display count equals a frequency of 1 Hertz (e.g., a display reading of
100 = 100 Hz.).
4.
If the
µ
-ECD frequency indicates contamination (
≥
1000) check for the
following:
•
Contaminated carrier gas trap(s) and or supply—replace carrier gas
supply tank and any traps on the carrier supply line.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......