Agilent 6890 Gas Chromatograph Service Manual
Micro-cell Electron Capture Detector (
µ
-ECD)
341
Replacement procedures
Replacing the makeup gas adapter
After removing the
µ
-ECD detector from the GC, you can further disassemble
it to replace the makeup gas adapter. The makeup gas adapter consists of a
line from the detector pneumatics manifold that carries makeup gas to a
weldment that screws into the bottom of the
µ
-ECD detector. From there, the
makeup gas sweeps past the end of the column and carries the column effluent
into the
µ
-ECD cell.
1.
Remove the detector as described in the Replacing the entire detector/
detector cell procedure in this section.
2.
Remove the Torx T-20 screw holding the pneumatics block(s) to the
detector manifold.
Note
There are two pneumatics blocks on the EPC version of the
µ
-ECD pneumatics
manifold. The outside block is the anode purge gas line and the inside block
is the make-up gas line.
3.
Slide the makeup gas adapter up and out of the GC.
4.
When re-installing the makeup gas adapter, ensure the following:
•
Approximately 6 inches of the makeup gas line resides in the oven
after installation.
•
The makeup gas line is bent into a coil or loop (inside the oven) that
loops around the bottom of the detector weldment and makeup gas
adapter.
Содержание 6890 Series
Страница 1: ...6890 Series Gas Chromatograph Service Manual ...
Страница 2: ...6890 Series Gas Chromatograph Service Manual ...
Страница 27: ...24 of 24 Jun 2001 Contents Contents Agilent 6890 Gas Chromatograph Service Manual ...
Страница 29: ......
Страница 39: ......
Страница 195: ...260 Solvent Vapor Exit Accessory Leak testing 10 of 10 Jun 2001 Inlets Agilent 6890 Gas Chromatograph Service Manual ...
Страница 207: ......
Страница 389: ...360 Auxiliary EPC Replacement procedures 16 of 16 Jun 2001 Detectors Agilent 6890 Gas Chromatograph Service Manual ...
Страница 391: ......
Страница 459: ......
Страница 479: ...520 Gases Supplying valve actuator air 14 of 14 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 485: ...530 Power Grounding 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 491: ...540 Instrument Specifications Detectors 6 of 6 Jun 2001 Site Preparation Agilent 6890 Gas Chromatograph Service Manual ...
Страница 492: ...600 Performance Verification How to check the performance of the 6890 instrument 610 Checkout Chromatograms ...
Страница 493: ......
Страница 521: ......
Страница 523: ......
Страница 555: ...840 Shutdowns 10 of 10 Jun 2001 Error Codes and Internal Diagnostics Agilent 6890 Gas Chromatograph Service Manual ...
Страница 580: ...900 Remote Access Remote Access information was not available for this printing ...
Страница 581: ......
Страница 583: ......
Страница 631: ......
Страница 637: ...1110 Valves Liquid sample valves 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 647: ...1120 Valve Box Removing the valve box assembly 10 of 10 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 675: ...1150 Troubleshooting Pressure check 6 of 6 Jun 2001 Valves Agilent 6890 Gas Chromatograph Service Manual ...
Страница 677: ......
Страница 717: ...1225 Main Board 6890N Fuses 6 of 6 Jun 2001 Electrical Agilent 6890 Gas Chromatograph Service Manual ...
Страница 737: ......