    ADMINISTRATIVE INFORMATION  

 

SERVICE NOTE CLASSIFICATION:

 

MODIFICATION RECOMMENDED

 

ACTION 

[[]]  ON SPECIFIED FAILURE 

CATEGORY: 

X  AGREEABLE TIME                                     

STANDARDS 
 

LABOR:  1.5 Hours 

                                    

LOCATION 

x  CUSTOMER INSTALLABLE 

CATEGORY: 

x  ON-SITE 

 

x  SERVICE CENTER 

 

[[]]  CHANNEL PARTNER 

SERVICE 

[[]]  RETURN 

INVENTORY: [[]]  SCRAP 
 

X  SEE TEXT 

USED 

[[]]  RETURN 

PARTS: 

 x    SCRAP 

 

[[]]  SEE TEXT 

AVAILABILITY: 

PRODUCT’S SUPPORT LIFE

 

NO CHARGE AVAILABLE UNTIL:  (Always) 

AUTHOR:   CP

 

PRODUCT LINE:  SP 

ADDITIONAL INFORMATION: This Service Note was originally posted as a Modification Available Service Note. 

   

© AGILENT TECHNOLOGIES, INC. 2008

 

    PRINTED IN U.S.A.   

  November 3,  2008

 

 

 

 

Rev. 15

 

Page 1 of 5

 

66103A-08A

  

S    E    R    V    I    C    E                N    O    T    E    

 

 

 

 

 

 

Supersedes: 

66103A-08   

66103A DC Power Supply 

 
Serial Numbers:  Serial Numbers:  MY40005783/MY40006268 
 

The instrument will not communicate or will not complete self test.   

Parts Required:  
 
P/N 

Description 

Qty. 

1200-1274      

SOCKET-IC- 68-CONT PLCC                           1

 

 

 

 

 

 

      1821-1479          IC 16-Bit MCU 16 MHZ CMOS  68PLCC      1                  

                                                   

 

   

6610X-61022     

PCA-TSTD PWRMESH                                      

 

 
IMPORTANT 

Consult the Service Manual for the 6610XA Power Modules p/n 5959-3364. The information in the Service Manual should be 
used during the disassembly and reassembly see page 57. The Service Manual can be viewed on line at the web address described 
below. 

 
When doing this Modification the following steps are required. 

1.

 

Follow proper procedures to eliminate ESD damage to the instrument and components. 

2.

 

The AC line MUST be turned OFF at the source and disconnected from the instrument. 

       3.     The correct hand tools should be used for disassembly and assembly procedures.   
 

http://cp.literature.agilent.com/litweb/pdf/5959-3364.pdf

 

Содержание 66102A

Страница 1: ... E N O T E Supersedes 66103A 08 66103A DC Power Supply Serial Numbers Serial Numbers MY40005783 MY40006268 The instrument will not communicate or will not complete self test Parts Required P N Description Qty 1200 1274 SOCKET IC 68 CONT PLCC 1 1821 1479 IC 16 Bit MCU 16 MHZ CMOS 68PLCC 1 6610X 61022 PCA TSTD PWRMESH IMPORTANT Consult the Service Manual for the 6610XA Power Modules p n 5959 3364 Th...

Страница 2: ...ESH p n 6610X 61022 should only be replaced if the traces on the printed circuit board are damaged during the removal of the 68 Pin PLCC socket p n 1200 1274 Circuit Reference U001 It is very unlikely that the PCA Tested assembly p n 6610X 61022 will need to be replaced if the 68 pin PLCC socket p n 1200 1274 is removed and replace with care Situation When the 16 Bit processor p n1821 1474 is inst...

Страница 3: ...viewed these markings should be used to identify a Good or Defective socket Socket A is a sample of a GOOD AMP socket Socket B is a sample of a GOOD Tyco Augit socket Socket C is a sample of a GOOD FCI socket Socket D is a sample of a DEFECTIVE Tyco socket The sockets above labeled A B or C are good and should NOT be replaced The socket above labeled D is considered defective should replaced The i...

Страница 4: ...nt was tested by comparing the mating and un mating forces on the older and newer designs The test results for sockets with the increased spring height newer design showed an increase in the mean mating force of 18 9 and an increase in the mean un mating force of 23 4 The increase in the un mating force test results between the old design and the new design indicate that a greater normal force is ...

Страница 5: ...5 66106A MY40000389 MY40000589 Testing Calibration NOTE When this modification is complete the instrument should be turned on using the 66000A mainframe The 6610XA should complete self test The instrument should also be programmed for voltage and current Complete performance testing is NOT required The instrument should output the programmed voltage and current and should pass the self test Changi...

Отзывы: