CI Troubleshooting
3
79
No or low PFDTD signal,
but reagent ions are normal
You are using any reagent gas but methane in PCI.
Switch to methane.
Wrong or corrupted tune file loaded
Check your tune file.
No PFDTD in the calibrant vial
Inspect the calibration vial on the back of the flow controller. If necessary, fill
the vial with PFDTD. Do not fill the vial completely; keep the level at least
0.5 cm from the top of the vial.
The pressure of the methane entering the flow controller is too high.
Ensure the regulator on the methane supply is set to 10 psig (70 kPa).
The CI ion source is dirty.
Clean the ion source.
The calibration valve was not purged after the vial was refilled.
Purge the calibration valve as described in
“To Purge the Calibration Valves”
on page 130. Then clean the ion source.
The calibrant vial was overfilled. Excess PFDTD can quench the chemical
ionization reactions.
Check the level of the PFDTD in the calibration vial. It should be
below
the end
of the inside tube in the vial.
Poor quality methane (purity below 99.99%)
Replace the methane with high-purity methane. If necessary, clean and purge
the reagent gas lines and clean the ion source.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...