64
3
CI Troubleshooting
Common CI-Specific Problems
Because of the added complexity of the parts required for CI, there are many
potential problems added. The greatest number of, and most serious problems
with CI are associated with leaks or contamination in the reagent gas
introduction system. NCI is especially sensitive to the presence of air; small
leaks that cause no problems in PCI can destroy NCI sensitivity.
As with EI, if the MSD tunes well and no air leak is present, sample sensitivity
problems should be addressed by GC inlet maintenance first.
•
Wrong reagent gas
•
Reagent gas not hooked up or hooked up to wrong reagent gas inlet port
•
Wrong ions entered in tune file
•
Wrong tune file selected
•
Not enough bakeout time has elapsed since vent (background is too high)
•
Wrong column positioning (extending > 2 mm past tip of interface)
•
Interface tip seal not installed
•
EI ion source installed in CI mode
•
EI filament or other EI ion source parts in CI ion source
•
Air leaks in reagent gas flow path
•
CI filament has stretched and sagged:
•
High EMV
•
Linear (no inflection point) electron energy (EIEnrgy) ramp
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...