Analyzer
7
5977B Series MSD Troubleshooting and Maintenance Manual
185
EI Ion Source
The EI ion source (See
on page 186) operates by electron ionization.
The sample enters the ion source from the GC/MSD interface. Electrons
emitted by a filament enter the ionization chamber, guided by a magnetic field.
The high-energy electrons interact with the sample molecules, ionizing and
fragmenting them. The positive voltage on the repeller pushes the positive ions
into the lens stack, where they pass through several electrostatic lenses. These
lenses concentrate the ions into a tight beam, which is directed into the mass
filter.
Ion source body
The ion source body (See
on page 186 and
a cylinder. It holds the other parts of the ion source, including the lens stack.
With the repeller, and in the SST/Inert ion source, the drawout plate, it forms
the ionization chamber. The ionization chamber is the space where the ions
are formed. Slots in the source body help the vacuum system to pump away
carrier gas and unionized sample molecules or fragments.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...