5977B Series MSD Troubleshooting and Maintenance Manual
6
Vacuum System
Overview
The vacuum system creates the high vacuum (low pressure) required for the
MSD to operate. Without the vacuum, the molecular mean free path would be
very short and ions would collide with air molecules before they could reach
the detector. Operation at high pressures also would damage analyzer
components.
The 5977B Series MSDs use two vacuum pumps to obtain the vacuum levels
needed. One of two foreline pumps (standard or dry) creates a low vacuum,
then a high vacuum pump engages to create the vacuum needed for
operation.The 5977B Series MSD uses one of two kinds of high vacuum pumps:
a diffusion pump or a turbomolecular (turbo) pump. The pump type
determines the maximum column flow supported by the MSD. The 5977B HES
Series MSD model G7079B uses a turbomolecular (turbo) pump for high
vacuum. It has a maximum column flow rate of 4.0 mL/min.
Most vacuum system operation is automated. Operator interaction is through
the data system or control panel. Monitor the vacuum system through the data
system or GC control panel.
Table 10
Recommended maximum flow rates by model
Model number
Description
Maximum recommended
column flow
G7080B
Diffusion pump, EI
1.5 mL/min
G7077B
Turbo pump, EI
4.0 mL/min
G7078B
Turbo pump, EI
4.0 mL/min
G7081B
Turbo pump, EI
4.0 mL/min
G7079B
Turbo pump, EI HES
4.0 mL/min
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...