![Agilent Technologies 5977B Series Скачать руководство пользователя страница 141](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608141.webp)
General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
141
Maintaining the Electronics
Scheduled maintenance
None of the electronic components of the MSD need to be replaced on a
regular schedule. None of the electronic components in the MSD need to be
adjusted or calibrated on a regular schedule. Avoid unnecessary handling of
the MSD electronics.
Electronic components
Very few of the electronic components are operator serviceable. The primary
fuses can be replaced by the operator. The RF coils can be adjusted by the
operator. All other maintenance of the electronics should be performed by
your Agilent Technologies service representative.
Electrostatic precautions
All of the printed circuit boards in the MSD contain components that can be
damaged by electrostatic discharge (ESD). Do not handle or touch these
boards unless absolutely necessary. In addition, wires, contacts, and cables
can conduct ESD to the printed circuit boards to which they are connected.
This is especially true of the mass filter (quadrupole) contact wires which can
carry ESD to sensitive components on the side board. ESD damage may not
cause immediate failure, but it will gradually degrade the performance and
stability of your MSD.
When you work on or near printed circuit boards, or when you work on
components with wires, contacts, or cables connected to printed circuit
boards, always use a grounded antistatic wrist strap and take other antistatic
WA R N I N G
Improper use of these procedures could create a serious safety hazard. Improper use
of these procedures could also result in serious damage to, or incorrect operation of,
the MSD.
WA R N I N G
Vent the MSD and disconnect its power cord before performing any of these
procedures
except
adjusting the RF coils.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...