5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Install the EI Calibration and Vent Valve Assembly
Materials needed
•
Calibration valve
•
Diffusion (G7077-60211)
•
Turbo (G7077-60204)
•
O-ring for calibration valve (0905-1217)
•
PFTBA (05971-60571) or other tuning compound
•
Screwdriver, Torx T-20 (8710-1615)
Procedure
1
Remove the old valve assembly. (See
“To Remove the EI Calibration and
2
Ensure the valve O-ring is in place. If it is worn or damaged, replace it.
3
Install the calibration and vent valve assembly, and tighten the screws that
hold it in place.
4
Reconnect the calibration valve cable to the connector next to the fan.
5
Remove the vial from the new calibration valve. (See
on page 128.) The valve is supplied with a vial already
installed.
6
Fill and reinstall the calibration vial.
7
Pump down the MSD.
C A U T I O N
Failure to purge the calibration valve will damage the filaments and detector.
Содержание 5977B Series
Страница 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Страница 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Страница 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Страница 88: ...88 3 CI Troubleshooting ...
Страница 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Страница 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Страница 196: ...196 5977B Series MSD Troubleshooting and Maintenance Manual 7 Analyzer Figure 46 Interface tip seal Interface tip seal ...
Страница 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Страница 232: ...232 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 55 Pfeiffer DUO pump ...
Страница 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Страница 243: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 243 Figure 63 Extractor EI ion source ...
Страница 245: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 245 Figure 64 The CI ion source 15 16 ...
Страница 250: ...250 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 67 CI GC MSD interface ...
Страница 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Страница 257: ......
Страница 258: ... G3870 90003 G7077 90028 Agilent Technologies Inc Printed in USA September 2016 Agilent Technologies ...