Materials needed
•
Carrier gas, (99.9995% pure or better)
•
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Procedure
1
Install the column in the GC inlet. (See
“To Install a Capillary Column in a
2
Set a minimum velocity of 30 cm/s, or as recommended by the column
manufacturer. Allow the carrier gas to flow through the column at room
temperature for 15 to 30 minutes to remove air.
3
Program the oven from room temperature to the maximum temperature
limit for the column.
4
Increase the temperature at a rate of 10 to 15 °C/min.
5
Hold at the maximum temperature for 30 minutes.
6
Set the GC oven temperature to 30 °C, and wait for the GC to become ready.
7
Attach the column to the GC/MSD interface. (See
Column in the GC/MSD Interface Using the Self-Tightening Column Nut”
page 42.)
WARNING
Do not condition your capillary column with hydrogen. Hydrogen accumulation
in the GC oven can result in an explosion. If you plan to use hydrogen as your
carrier gas, first condition the column with ultrapure (99.999% or better) inert
gas such as helium, nitrogen, or argon.
WARNING
The GC operates under high temperatures. Do not touch any GC parts unless
you are certain they are cool.
CAUTION
Never exceed the maximum column temperature, either in the GC/MSD
interface, the GC oven, or the inlet.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...