To Replace a Filament in an EI XTR, SS, or
Inert Source
Materials needed
•
Filament assembly (G7005-60061)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5-mm (8710-1570)
Procedure
1
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
3
Remove the ion source. (See
“To Remove the EI XTR, SS, or Inert Source”
4
Remove the gold plated screw and washer from the filament. (See
5
Slide the filament assembly off the ion source assembly.
WARNING
The analyzer operates at high temperatures. Do not touch any part until you
are sure it is cool.