Operating in EI Mode
To Verify EI System Performance
5977B Series MSD Operating Manual
131
To Verify EI System Performance
Materials needed for an EI HES
•
10 fg/µL OFN Checkout standard 3 × 1 mL (5190-0585)
Materials needed for an EI XTR, SS, or Inert source
•
1 pg/µL OFN EI Checkout standard 3 × 1 mL (5188-5348)
Verify the tune performance
1
Verify that the system has been pumping down for at least 60 minutes.
2
Set the GC oven temperature to 150 °C and the column flow to 1.0 mL/min.
3
In the
Instrument Control
view, select
Checkout Tune
from the
Checkout
menu. The software will perform an autotune and print the report.
4
When the autotune has completed, save the method, and then select
Evaluate Tune
from the
Checkout
menu.
The software will evaluate the last autotune, and print a System Verification
– Tune report.
Verify the sensitivity performance
1
Set up to inject 1 µL of the checkout sample shown in the Materials needed
for your source type above, either with the ALS or manually.
2
In the
Instrument Control
view, select
Sensitivity Check
from the
Checkout
menu. The system displays an
Alert
dialog reminder about resolving the
OFN_SN method and placing the OFN sample in vial 1 when an ALS is
configured.
3
If necessary, resolve your hardware with this method and place the sample in
the vial 1 position.
4
Click
OK
to run the method.
When the method is completed, an evaluation report will be printed.
Verify that rms signal-to-noise ratio meets the published specification. See the
Agilent Web site at www.agilent.com/chem for specifications.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...