Chromatographic Symptoms
3
Troubleshooting and Maintenance Guide
67
•
Source temperature sensor is shorted
•
Mass filter (quad) temperature sensor is open
•
Mass filter (quad) temperature sensor is shorted
•
No heater voltage (heater fuse has probably failed)
•
Heater voltage is too low
•
Temperature zone has timed out (heater failed, bad heater
wiring, or loose temperature sensor)
•
Problem with the temperature control electronics
•
Source heater is open
•
Source heater is shorted
•
Mass filter heater is open
•
Mass filter heater is shorted
The high vacuum pump is not ready
•
Turbo pump is on but has not had enough time (5 minutes)
to reach 80% of its normal operating speed
•
Turbo pump is not working correctly
•
MSD electronics are not working correctly
The system is in standby
This message is triggered by a shutdown signal on the remote
start cable. It is usually caused by a GC fault, an ALS fault, or a
bad cable connection. Once the cause of the fault is corrected,
selecting
MS ON
or checking MSD status should remove the
message.
The system is in vent state
The message says the system is vented, but if the fault has just
occurred it may still be under vacuum and the turbo pump may
still be at high speed. Wait at least 30 minutes after seeing this
message before you actually vent the MSD.
•
System was vented on purpose (no problem)
C A U T I O N
Venting the MSD too soon after this message appears can damage
a turbo pump.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...