Temperature Symptoms
The instrument has the following heated zones:
•
Ion source (
Source
in the MSD ChemStation software)
•
Mass filter (
Quad
in the MSD ChemStation software)
•
GC/MSD interface
Thermal Aux #2
in the MSD ChemStation
software
•
LTM column module
•
Heated guard column enclosure
•
Inlet
Each heated zone has a heater and temperature sensor.
Ion source will not heat up
•
High vacuum pump is off or has not reached normal
operating conditions*
•
Incorrect temperature setpoint
•
Ion source has not had enough time to reach temperature
setpoint
•
Ion source heater cartridge is not connected*
•
Ion source temperature sensor is not connected*
•
Ion source heater failed (burned out or shorted to ground)*
•
Ion source temperature sensor failed*
•
Source power cable is not connected to the side board*
•
MSD electronics are not working correctly
* These will cause an error message.
Mass filter (quad) heater will not heat up
•
High vacuum pump is off or has not reached normal
operating conditions*
•
Incorrect temperature setpoint
•
Mass filter has not had enough time to reach temperature
setpoint
•
Mass filter heater cartridge is not connected*
•
Mass filter temperature sensor is not connected*
•
Mass filter heater failed (burned out or shorted to ground)*
•
Mass filter temperature sensor failed*
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...