Troubleshooting and Maintenance Guide
3
Chromatographic Symptoms
Foreline pressure is too low
If the pressures you observe are below 20 m Torr, check for the
following:
•
Column (carrier gas) flow is too low
•
Column plugged or crushed by an overtightened nut
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
•
Foreline gauge is not working correctly
* These could create a fault condition in the GC that would prevent
the GC from operating.
Analyzer chamber pressure is too low
If the pressures you observe are below 1 × 10
-6
Torr, check for
the following:
•
Column (carrier gas) flow is too low
•
Column plugged or crushed by overtightened nut
•
Empty or insufficient carrier gas supply*
•
Bent or pinched carrier gas tubing*
* These could create a fault condition in the GC that would prevent
the GC from operating.
Gauge controller displays 9.9+9 and then goes blank
This indicates the pressure in the analyzer chamber is above 8 ×
10
-3
Torr.
•
Solvent peak from an on-column injection
•
MSD has not had enough time to pump down
•
Excessive foreline pressure
•
Vacuum gauge has failed
•
Line voltage too low
•
Turbo pump is not working correctly
Power indicator on the gauge controller does not light
•
Unplugged gauge controller power cord
•
Incorrect or inadequate line voltage (24 V supply)
•
Failed gauge controller fuse
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...