Analyzer
13
Troubleshooting and Maintenance Guide
203
Filaments
Two filaments (
) are located on opposite sides of the
outside of the EI ion source. The
active
filament carries an
adjustable AC emission current. The emission current heats the
filament causing it to emit electrons which ionize the sample
molecules. In addition,
both
filaments have an adjustable DC
bias voltage. The bias voltage determines the energy on the
electrons, usually –70 eV.
The filament is shut off automatically if there is a general
instrument shutdown. Three parameters affect the filaments:
filament selection (Filament), filament emission (Emission)
current, and electron energy (EIEnrgy).
Filament selection
The filament selection parameter (Filament) selects which
filament in the ion source is active.
Sometimes, one EI filament will give better performance than
the other. To select the better of the two filaments, run two
autotunes, one with each filament. Use the filament that gives
the best results.
Emission current
The filament emission current (Emission) is variable between 0
and –315 µA, but should be set to the software default for
normal operation.
Electron energy
The electron energy (EIEnrgy) is the amount of energy on the
ionizing electrons. It is determined by the bias voltage; –70 VDC
bias on the filament causes emitted electrons to possess –70 eV
(electron volts). This value is adjustable from –5 to –241 VDC,
but for normal operation, set this parameter to –70.
Filament care
Like the filaments in incandescent light bulbs, the ion source
filaments will eventually burn out. Certain practices will reduce
the chance of early failure:
•
If you have an optional G3397A Micro-Ion vacuum gauge,
use it to verify that the system has an adequate vacuum
before turning on the analyzer, especially after any
maintenance was performed.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...