Side Plate
The side plate (
) covers the large opening in the side of
the analyzer chamber. It is attached to the manifold with a
hinge. The analyzer assembly is attached to the side plate inside
the analyzer chamber. The hinge allows the side plate to swing
away from the manifold for easy access to the analyzer.
Several electrical feedthroughs are built into the side plate.
Wires connect the feedthroughs to analyzer components. The
electronic side board is mounted on the atmospheric side of the
side plate.
Thumbscrews are located at each end of the side plate.
C A U T I O N
Fasten both side plate thumbscrews for shipping or storage only. For
normal operation, both thumbscrews should be loose. For operation
with hydrogen carrier gas, or with flammable or explosive CI reagent
gases, the front thumbscrew should be fastened just finger-tight.
Overtightening will warp the side plate and cause air leaks. Do not
use a tool to tighten the side plate thumbscrews.
C A U T I O N
When you turn on the power to pump down the MSD, be sure to
press on the side board to ensure a good seal.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...