General Maintenance
10
Troubleshooting and Maintenance Guide
155
To Remove the Heater and Sensor from the Mass Filter
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
Procedure
1
Vent the instrument.
2
Open the analyzer chamber.
3
Disconnect the mass filter heater and temperature sensor
wires from the feedthrough board.
4
Remove the mass filter heater assembly from the mass filter
radiator.
C A U T I O N
Do not touch the mass filter contact leads. This could cause ESD
damage to the side board.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...