To Lubricate the Side Plate O-Ring
Materials needed
•
Cloths, clean (05980-60051)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Grease, Apiezon L, high vacuum (6040-0289)
The side plate O-ring needs a thin coat of grease to ensure a
good vacuum seal. If the side plate O-ring appears dry or does
not seal correctly, lubricate it using this procedure. A good test
is to wipe off the side plate with methanol, then close the
analyzer chamber. If the O-ring has enough grease on it, it will
leave a faint trace on the side plate.
Procedure
1
Vent the instrument.
2
Open the analyzer chamber. (See the
5975T LTM GC/MSD
Operation Manual
).
3
Remove the analyzer by disconnecting the source wiring,
and loosening the sideboard from the hinge. (See the
5975T
LTM GC/MSD Operation Manual
).
4
Use a clean, lint-free cloth or glove to spread a
thin
coat of
high vacuum grease only on the exposed surface of the
O-ring (
).
5
Use a clean, lint-free cloth or glove to wipe away excess
grease. If the O-ring looks shiny, there is too much grease on
it.
C A U T I O N
Vacuum seals other than the side plate O-ring and vent valve O-ring
do not need to be lubricated. Lubricating other seals can interfere
with their correct function.
C A U T I O N
Do not use anything except the recommended vacuum grease.
Excess grease can trap air and dirt. Grease on surfaces of the O-ring
other than the exposed surface can trap air, resulting in air spikes
during operation.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...