General Maintenance
10
Troubleshooting and Maintenance Guide
129
To Reinstall the EI Calibration and Vent Valve Assembly
Materials needed
•
Calibration valve (G3170-60204)
•
O-ring for calibration valve (0905-1217)
•
PFTBA (05971-60571) or other tuning compound
•
Screwdriver, Torx T-15 (8710-1622)
Procedure
1
Remove the old valve assembly. See
2
Be sure the valve O-ring is in place. If it is worn or damaged,
replace it.
3
Install the calibration and vent valve assembly and tighten
the screws that hold it in place.
4
Reconnect the calibration valve cable to the connector next
to the fan.
5
Remove the vial from the new calibration valve. See
. The valve is supplied with a vial already installed.
6
Fill and reinstall the calibration vial. See
.
7
Pump down the MSD.
8
Select
Purge Calibrant Valve
from the
Vacuum
menu in the
Tune
and Vacuum Control
view.
C A U T I O N
Failure to purge the calibration valve will damage the filaments and
detector.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...