![Agilent Technologies 5975T LTM GC/MSD Скачать руководство пользователя страница 100](http://html1.mh-extra.com/html/agilent-technologies/5975t-ltm-gc-msd/5975t-ltm-gc-msd_troubleshooting-and-maintenance-manual_2867607100.webp)
100
Troubleshooting and Maintenance Guide
8
Checking for Leaks
Leak Check Tips
When checking for leaks, consider the system in two parts:
external leak points and instrument leak points.
•
External leak points
include the gas cylinder (or gas
purifier), regulator and its fittings, supply shutoff valves,
and connections to the GC supply fittings.
•
Instrument leak points
include inlets, LTM column
connections, GC/MSD transfer line connection, and
connections between flow modules and inlets.
1
Gather the following:
•
Electronic leak detector capable of detecting the gas type
•
7/16-in, 9/16-in, and 1/4-in wrenches for tightening
Swagelok and column fittings
2
Check any potential leak points associated with any
maintenance recently performed.
3
Check instrument capillary fittings and connections that
undergo thermal cycling, since thermal cycling tends to
loosen some fitting types. Use the electronic leak detector to
determine if a fitting is leaking.
•
Start by checking any newly made connections first.
•
Remember to check connections in the gas supply lines
after changing traps or supply cylinders.
WA R N I N G
Hydrogen (H
2
) is flammable and is an explosion hazard when
mixed with air in an enclosed space (for example, a flow meter).
Purge flowmeters with inert gas as needed. Always measure
gases individually.
WA R N I N G
Hazardous sample gases may be present.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...