![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 205](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605205.webp)
Analyzer
7
5975 Series MSD Troubleshooting and Maintenance Manual
205
Filament care
Like the filaments in incandescent light bulbs, the ion source filaments will
eventually burn out. Certain practices will reduce the chance of early failure:
•
If you have an optional G3397A Micro-Ion vacuum gauge, use it to verify
that the system has an adequate vacuum before turning on the analyzer,
especially after any maintenance was performed.
•
If you are controlling your MSD from the Edit Parameters screen, always
select MSOff before changing any of the filament parameters.
•
When setting up data acquisition parameters, set the solvent delay so that
the analyzer will
not
turn on while the solvent peak is eluting.
•
When the software prompts Override solvent delay? at the beginning of a
run,
always
select NO.
•
Higher emission current will reduce filament life.
•
Higher electron energy will reduce filament life.
•
Leaving the filament on for short times (
1
minute) during data acquisition
will reduce filament life.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...