![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 164](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605164.webp)
164
5975 Series MSD Troubleshooting and Maintenance Manual
5
CI Maintenance
To Clean the CI Ion Source
The CI ion source has slightly different cleaning requirements than the EI ion
source. See the video clip in the 5975 Series MSD User Information DVD.
Frequency of cleaning
Because the CI ion source operates at much higher pressures than the EI ion
source, it will probably require more frequent cleaning than the EI ion source.
Cleaning of the source is not a scheduled maintenance procedure. The source
should be cleaned whenever there are performance anomalies that are
associated with a dirty ion source. See
for
symptoms that indicate a dirty ion source.
Visual appearance is not an accurate guide to cleanliness of the CI ion
source. The CI ion source can show little or no discoloration yet still need
cleaning.
Let analytical performance be your guide.
Procedure
Cleaning the CI ion source (
) is very similar to cleaning the EI ion
source. Use the cleaning procedure in
135 with the following exceptions:
•
The CI ion source may not look dirty but deposits left by chemical
ionization are very difficult to remove. Clean the CI ion source thoroughly.
•
Use a round wooden toothpick to gently clean out the electron entrance
hole in the source body and the ion exit hole in the drawout plate.
•
Do not use halogenated solvents. Use hexane for the final rinse.
C A U T I O N
Do not use halogenated solvents to clean the CI ion source.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...