Triggering the Oscilloscope
To use Duration triggering
To use Duration triggering
Duration trigger lets you define a pattern, then trigger on a specified time
duration of this logical AND combination of the channels.
1 Press the
More
key in the Trigger section of the front panel, rotate the
Entry knob until
Duration
is displayed in the
Trigger
softkey, then press
the
Settings
softkey to display the Duration trigger menu.
2 For each analog or digital channel you want to include in the desired
pattern, press the
Channel
softkey to select the channel.
This is the channel source for the H, L, or X condition. As you press the
Channel
softkey (or rotate the Entry knob on mixed-signal oscilloscopes), the channel
you select is highlighted in the
Pattern =
line directly above the softkeys and
in the upper-right corner of the display next to "
Dur
". External trigger may also
be specified as a channel in the pattern when using the 2-channel and 4-channel
oscilloscopes.
Adjust the trigger level for the selected analog channel by turning the Trigger
Level
knob. The position of the trigger level for the analog channel is indicated
by the trigger level icon
at the far left side of the display when DC coupling
is selected Set the threshold for the selected digital channel group by pressing
the
D7 Thru D0
or
D15 Thru D8
key then press the
Threshold
softkey. The value
of the trigger level or digital threshold is displayed in the upper-right corner of
the display.
Trigger level
or threshold
Selected
channel
Duration
trigger
Channel
select
Logic level
Return to
previous menu
Qualifier
Qualifier time set
T
Содержание 54621A
Страница 4: ...iv ...
Страница 10: ...Contents 6 ...
Страница 11: ...1 Getting Started ...
Страница 38: ...1 28 ...
Страница 39: ...2 Front Panel Overview ...
Страница 63: ...3 Triggering the Oscilloscope ...
Страница 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 171: ...5 Making Measurements ...
Страница 186: ...5 16 Making Measurements To use the XY horizontal mode Signals are 90 out of phase Signals are in phase ...
Страница 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Страница 223: ...6 Utilities ...
Страница 239: ...7 Performance Characteristics ...
Страница 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Страница 260: ...7 22 ...