5- 18
Making Measurements
Math Scale and Offset
Math Scale and Offset
Any math function may be manually scaled by pressing the
Settings
softkey, and
then adjusting the Scale or Offset value.
1 Press the
Settings
softkey in the Math menu to set your own scale factors
(units/division) or offset (units) for the selected math function.
Units for each input channel can be set to Volts or Amps using the channel
Probe
Units
softkey. Scale and offset units are:
A scale unit of
U
(undefined) will be displayed for math function 1-2 and for
d/dt, and
∫
dt when 1-2 or 1+2 is the selected source if channel 1 and channel 2
are set to dissimilar units in the channel
Probe Units
softkey.
2 Press the
Scale
or
Offset
softkey, then turn the Entry knob to rescale or
change the offset value for your math function.
Math Scale and Offset are Set Automatically
Any time the currently displayed math function definition is changed, the function is
automatically scaled for optimum vertical scale and offset. If you manually set scale
and offset for a function, select a new function, then select the original function, the
original function will be automatically rescaled.
Math function
Units
FFT
dB
*
(decibels)
1*2
V
2
, A
2
, or W (Volt-Amp)
1-2
V or A
d/dt
V/s or A/s (V/second or A/second)
∫
dt
Vs or As (V-seconds or A-seconds)
*
When the FFT source is channel 1 or channel 2 (or channel 3 or 4 on the 54624A), FFT units
will be displayed in dBV when channel units is set to Volts and channel impedance is set to
1 M
Ω
. FFT units will be displayed in dBm when channel units is set to Volts and channel
impedance is set to 50
Ω
(54640-series only). FFT units will be displayed as dB for all other FFT
sources or when a source channel’s units has been set to Amps.
Содержание 54621A
Страница 4: ...iv ...
Страница 10: ...Contents 6 ...
Страница 11: ...1 Getting Started ...
Страница 38: ...1 28 ...
Страница 39: ...2 Front Panel Overview ...
Страница 63: ...3 Triggering the Oscilloscope ...
Страница 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 171: ...5 Making Measurements ...
Страница 186: ...5 16 Making Measurements To use the XY horizontal mode Signals are 90 out of phase Signals are in phase ...
Страница 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Страница 223: ...6 Utilities ...
Страница 239: ...7 Performance Characteristics ...
Страница 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Страница 260: ...7 22 ...