3- 52
To use USB triggering
USB trigger will trigger on a Start of Packet (SOP), End of Packet (EOP) signal,
Reset Complete (RC), Enter Suspend (Suspend), or Exit Suspend (Exit Sus)
on the differential USB data lines (D+ and D-). USB Low Speed and Full Speed
are supported by this trigger.
1 Press the
More
key in the Trigger section of the front panel, rotate the
Entry knob until
USB
is displayed in the
Trigger
softkey, then press the
Settings
softkey to display USB trigger menu.
2 Press the
Speed
softkey to select the speed of the transaction being
probed.
You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s).
D+,–
D–,+
Bus Idle
SOP
SE0 portion of EOP
End of packet trigger
Signal
speed
D+ source
channel
D– source
channel
Trigger on:
condition
Trigger level
or threshold
Currently selected
data source
USB trigger
Return to
previous menu
Содержание 54621A
Страница 4: ...iv ...
Страница 10: ...Contents 6 ...
Страница 11: ...1 Getting Started ...
Страница 38: ...1 28 ...
Страница 39: ...2 Front Panel Overview ...
Страница 63: ...3 Triggering the Oscilloscope ...
Страница 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Страница 171: ...5 Making Measurements ...
Страница 186: ...5 16 Making Measurements To use the XY horizontal mode Signals are 90 out of phase Signals are in phase ...
Страница 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Страница 223: ...6 Utilities ...
Страница 239: ...7 Performance Characteristics ...
Страница 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Страница 260: ...7 22 ...