background image

8 Installation Note 5967-8554

16. Attach the cable from the rear-panel connector labeled GATE TRIGGER INPUT to A16J501. See

Figure 3

.

17. Attach the cable from the rear-panel connector labeled GATE TRIGGER OUTPUT to A16J500. See

Figure 3

.

18. Reattach I/O board to CPU. Be careful not to pinch the SMB cable.

19. If the instrument has the dual I/O option, then it should be reattached. Make sure that the

connector is properly aligned, then firmly push into place. Replace the five screws attaching the
dual I/O to the rear panel.

CAUTION 

Be careful not to bend or misalign the connector pins.

Replacing the Instrument Cover

CAUTION 

To prevent damage when replacing the instrument cover, place a soft cloth or
towel between the work surface and the front panel.

Ensure cables do not bind between the instrument cover and the analyzer internal
assemblies.

20. Carefully place the analyzer on the work surface with the front panel facing down.

21. Replace the instrument cover assembly. The seam of the cover should be on the bottom side of the

analyzer.

22. Tighten the four rear-feet screws with a 4-mm hex wrench.

23. Replace the four screws and washers attaching the instrument cover assembly to the rear frame.

Contacting Agilent Technologies

1. If the CPU board does not have J501 and J500, order the 08590-60361 CPU kit for a compatible

CPU board.

2. To complete the Option 105 upgrade process refer to the Web at http://mktwww.soco.agilent.com/

field/service/signal/8590/105revb.htm to receive a complete set of instructions or call (707) 577-6801.

3. For detailed information about the 8590 Series firmware history, look on the Web at http://

mktwww.soco.agilent.com/field/service/signal/8590/firmhist.htm. The two different firmware kits
are noted on this web page.

Содержание 5062-8218

Страница 1: ...Part Number 5967 8554 Supersedes 5962 5078 Printed in USA September 2000 Installation Note Time Gated Spectrum Analyzer Capability Installation Kit 5062 8218 ...

Страница 2: ...d to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Copyright Agilent Technologies Inc 2000 ...

Страница 3: ...00 9999A99999 0000U00000 9999U99999 To Be Performed By Agilent Technologies Service Center Only Estimated Installation Time 1 2 Hour Estimated Verification Time 1 Hour Item Quantity Description Part Number 1 2 Nut hex 15 32 32 0590 1251 2 1 Cable BNC m to m 8120 2682 3 1 Option 105 product note 5960 2555 4 2 Adapters BNC female to male right angle 1250 0076 5 1 Cable BNC f to SMB f gate in 08591 6...

Страница 4: ...wer switch OFF and unplug the analyzer Failure to unplug the analyzer can result in personal injury CAUTION Electrostatic discharge ESD can damage or destroy electronic components All work on electronic assemblies should be performed at a static safe work station Refer to the installation and verification manual for your spectrum analyzer for information about a static safe work station and static...

Страница 5: ...s attaching the dual I O to the rear panel 8 Remove dual I O connector board from the instrument by pulling up on the board 9 Remove three screws on each side of the rear panel and pull the rear panel gently away from the instrument a few centimeters 10 Remove the plugs from the Gate Trigger In and Gate Out holes on the analyzer rear panel Use one pair of long needle nose pliers to compress the ba...

Страница 6: ... rear panel GATE OUTPUT 2 opening Use a hex nut to attach the cable to the rear panel Tighten the hex nut with the hex wrench 13 Remove the 4 screws attaching the Dual I O interconnect board A40 to the CPU and remove the I O board 14 Route the SMB cable ends to enable them to attach to A16J501 and A16J500 NOTE If J501 and J500 do not exist on the CPU you will need to order a CPU Board Kit Refer to...

Страница 7: ...Installation Note 5967 8554 7 Figure 3 Attaching Cables to Board Assembly ...

Страница 8: ...t bind between the instrument cover and the analyzer internal assemblies 20 Carefully place the analyzer on the work surface with the front panel facing down 21 Replace the instrument cover assembly The seam of the cover should be on the bottom side of the analyzer 22 Tighten the four rear feet screws with a 4 mm hex wrench 23 Replace the four screws and washers attaching the instrument cover asse...

Отзывы: