62
Chapter 6
Reading Out Measured Result
Reading Out Measured Result
Reading out measured result using
command
This command actually performs two tasks: it triggers the instrument and returns the
results. It is useful, for example, when you want to retrieve measurement results
immediately after triggering the instrument from an external controller.
The readout procedure using the
command is described below.
Step 1.
Use the
command on page 196 to set up the trigger mode to the GPIB
trigger (Bus).
Step 2.
Execute the
command.
Step 3.
Read out the measured result. To repeat the measurement, repeat Steps 2 and 3.
Example 6-1 and Example 6-2 show sample programs using the
6-1 is for the ASCII transfer format and Example 6-2 is for the binary transfer format.
These programs are stored on the sample program disk under the filenames “trg_a.bas” and
“trg_b.bas.”
These programs read out the measurement status, measured primary parameter result,
measured secondary parameter result, and comparator sorting result (when the comparator
function is ON) and then display this information.
The program of Example 6-1 is detailed below.
Line 60
Sets the GPIB address.
Line 70
Sets the data transfer format to the ASCII format.
Lines 90 to 100
Turns ON the trigger system continuous startup and sets the trigger
mode to the GPIB trigger.
Lines 120 to 130 Reads out the ON/OFF state of the comparator function and assigns it
to the Comp_flag variable.
Lines 170 to 210 After the trigger system state transitions to the trigger wait state,
triggers the instrument.
Line 230
If the Comp_flag value is 1 (the comparator function is ON), reads out
the measurement status, measured primary parameter result, measured
secondary parameter result, and comparator sorting result.
Line 250
If the Comp_flag value is not 1 (the comparator function is OFF),
reads out the measurement status, measured primary parameter result,
and measured secondary parameter result.
Lines 300 to 320 Displays the measurement status, measured primary parameter result,
and measured secondary parameter result.
Line 340
If the Comp_flag value is 1 (the comparator function is ON), displays
the comparator sorting result.
Line 360
If the Comp_flag value is not 1 (the comparator function is OFF),
displays “Comparator : OFF.”
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...