274
Appendix G
4268A vs. 4288A GPIB Command Correspondence Table
4268A vs. 4288A GPIB Command Correspondence Table
4268A vs. 4288A GPIB Command Correspondence Table
Table G-1 (by function) and Table G-2 (by alphabetical order) provide at-a-glance lists of
the GPIB command correspondences between the 4268A and the 4288A.
Table G-1
At-a-glance GPIB command correspondence between 4268A and 4288A (by
function)
Function
Item to Be Set Up/Executed
GPIB command
Note
4268A
4288A
Measurement
condition
Reset
:SYST:PRES
←
The 4288A does not
reset compensation
data.
*RST
←
Measurement
parameter
setup
Primary parameter
:CALC1:FORM
←
Secondary parameter
:CALC2:FORM
←
Measurement
signal setup
Frequency
:SOUR:FREQ
←
1 MHz frequency shift
N/A
:SYST:FSH
Level
:SOUR:VOLT
←
Auto level control
(ALC) function
ON/OFF
:SOUR:VOLT:ALC
N/A
The 4288A does not
support the ALC
function.
Output mode
:SOUR:VOLT:MODE
N/A
The 4288A does not
support the synchronous
source function.
Measurement
range setup
Ranging (auto/manual)
:RANG:AUTO
←
Range
:RANG
←
Measurement time mode setup
:APER
←
Averaging
setup
ON/OFF
:AVER
←
Number of counts
:AVER:COUN
←
Cable length setup
:CAL:CABL
←
Source delay setup
:TRIG:DEL
N/A
The 4288A supports
only trigger delay.
Trigger delay setup
:TRIG:SEQ2:DEL
:TRIG:DEL
Trigger mode setup
:TRIG:SOUR
←
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...