Chapter 10
175
Command Reference
:DATA:POIN
1
0
. C
o
mm
and
Refe
re
nc
e
:DATA:POIN
Syntax
:DATA:POINts {BUF1|BUF2|BUF3},<numeric>
:DATA:POINts? {BUF1|BUF2|BUF3}
Description
Specifies the size of data buffer 1, data buffer 2, or data buffer 3 in number of
measurements. Executing this command returns the pointer to the specified data buffer (the
location to feed measurement data) back to the start.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
BUF1
Specifies data buffer 1 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 1.
BUF2
Specifies data buffer 2 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 2.
BUF3
Specifies data buffer 3 as the buffer to which the <numeric> setup
is applied or reads out the setup of data buffer 3.
<numeric>
Description
Number of measurements for the data buffer specified with
{BUF1|BUF2|BUF3}
Range
Data buffer 1: 1 to 200
Data buffer 2: 1 to 200
Data buffer 3: 1 to 1000
Initial value
Data buffer 1: 200
Data buffer 2: 200
Data buffer 3: 1000
Resolution
1
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...