152
Chapter 10
Command Reference
:CALC2:MATH:STAT
:CALC2:MATH:STAT
Syntax
:CALCulate2:MATH:STATe {ON|OFF|1|0}
:CALCulate2:MATH:STATe?
Description
Determines whether to use the function (deviation measurement mode) that displays the
secondary parameter measurement result as deviation from the reference value (set using
the
command).
NOTE
Changing any of the measurement parameters will automatically disable this function. To
avoid this, you have to write a program so that this command is executed after the
execution of the measurement parameter setup command (
and
Parameters
Query response
{1|0}<newline><^END>
Related commands
Equivalent key
sequence
[
∆
Mode]
(
[Blue]
-
[Meas Prmtr]
) -
Sec
Description
ON or 1
Enables the deviation measurement mode.
OFF or 0 (initial value)
Disables the deviation measurement mode (i.e., displays
the measurement result as an absolute value).
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...